MRI Safety Information 
The AccelXâ„¢ Implant System has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating, migration, or image artifact in the MR environment. The safety of AccelXâ„¢ Implant System in the MR environment is unknown. Scanning a patient who has this device may result in patient injury.

MR Conditional Warning: The RF safety of the device has not been tested. The patient may only be imaged by landmarking at least 30 cm from the implant, or ensuring the implant is located outside of the RF coil.

 
A patient with this device can be scanned safely in an MR system under the following conditions:

Device Name

Accelx Implant System


Static Magnetic Field Strength (B0)

≤ 3 .0 T

Maximum Spatial Field Gradient
20 T/m (2 ,000 gauss/cm)
RF Excitation
Circularly Polarized (CP)
RF Transmit Coil Type

For body transmit coil, landmarking at least 30 cm from the implant, or ensuring the implant is located outside of the coil.

Extremity T/R coils permitted. Excludes Head T/R coil.

Operating Mode

Normal Operating Mode in the allowed imaging zone

Maximum Whole-Body
SAR
2 W/kg (Normal Operating Mode)
Maximum Head SAR
Not evaluated f or head landmark
Scan Duration

No specific constraints due to implant heating

Part#IFUMRI Rev 00

Download MRI Safety information
MRI Safety Information 
The AccelXâ„¢ Implant System has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating, migration, or image artifact in the MR environment. The safety of AccelXâ„¢ Implant System in the MR environment is unknown. Scanning a patient who has this device may result in patient injury.

MR Conditional Warning: The RF safety of the device has not been tested. The patient may only be imaged by landmarking at least 30 cm from the implant, or ensuring the implant is located outside of the RF coil.

 
A patient with this device can be scanned safely in an MR system under the following conditions:

Device Name

Accelx Implant System


Static Magnetic Field Strength (B0)

≤ 3 .0 T

Maximum Spatial Field Gradient
20 T/m (2 ,000 gauss/cm)
RF Excitation
Circularly Polarized (CP)
RF Transmit Coil Type

For body transmit coil, landmarking at least 30 cm from the implant, or ensuring the implant is located outside of the coil.

Extremity T/R coils permitted. Excludes Head T/R coil.

Operating Mode

Normal Operating Mode in the allowed imaging zone

Maximum Whole-Body
SAR
2 W/kg (Normal Operating Mode)
Maximum Head SAR
Not evaluated f or head landmark
Scan Duration

No specific constraints due to implant heating

Part#IFUMRI Rev 00